SREL Reprint #2980
|
||
|
|
Cryoetch And Cryo-planing For Low Temperature HRSEM: SE-I Imaging Of Hydrated Multicellular, Microbial And Bioorganic Systems R. P. Apkarian1, S. A. Shamsi2, S. A. Rizvi2, G. Benian3, A. L. Neal4, J. V. Taylor1, and S. N. Dublin1 1Integrated
Microscopy & Microanalytical Facility, Emory University, Atlanta,
GA 30322 Introduction: Low temperature (LT) preparation of organic species and biological specimens in small volumes (< 5 µl) by plunging in liquefied ethane or high pressure freezing (HPF) ensures rapid cryoimmobilization for subsequent processing and imaging with cryo-HRSEM. Provided that specimen concentrations are high enough to act as its own cryoprotectant, hexagonal ice formation may be avoided and the sample is preserved in the low temperature state. Subsequent processing by cryogenic fracture, cryo-planing, and high vacuum sublimation (etching) relieve surfaces for structural analysis in the nanometer range. Successful LT processing is completed by application of a thin monoatomic chromium (Cr) film that enriches the SE-I contrast for high resolution cryo-HRSEM recordings. SREL Reprint #2980 Apkarian, R. P., S. A. Shamsi, S. A. Rizvi, G. Benian, A. L. Neal, J. V. Taylor, and S. N. Dublin. 2006. Cryoetch And Cryo-planing For Low Temperature HRSEM: SE-I Imaging Of Hydrated Multicellular, Microbial And Bioorganic Systems. Microscopy and Microanalysis 12(Supp 2): 1120-1121.
|
|